Sensors (Basel) · 2025 · Journal article
Reliability and validity of single-camera markerless motion capture for shoulder range of motion
Original title: Reliability and Validity of the Single-Camera Markerless Motion Capture System for Measuring Shoulder Range of Motion in Healthy Individuals and Patients with Adhesive Capsulitis: A Single-Center Study
Lee S, Lee U, Kim Y, Noh S, Lee H, Lee S
Related research areas:Musculoskeletal pain disorders3D motion analysis
Summary
- Research question
- Whether a single-camera markerless motion-capture system (iBalance) measures shoulder range of motion reliably.
- Methods
- 30 healthy adults and 10 patients with adhesive capsulitis; seven shoulder movements measured three times by two raters with iBalance and a goniometer, repeated after one week (single-center pilot).
- Key results and limitations
- Excellent inter- and intra-rater reliability for flexion, abduction and passive abduction (ICC 0.91–0.98) and strong validity against the goniometer (ICC 0.85–0.97). Patients showed more variability; no adverse events. Other movements need algorithm improvement.
Summarized from the abstract. Please read the original article alongside.
Abstract
Assessing shoulder joint range of motion (ROM) is essential for diagnosing musculoskeletal disorders and optimizing treatments. This single-center pilot study evaluated the reliability and validity of iBalance, a single-camera markerless motion capture system, for measuring shoulder ROM. Forty participants (30 healthy individuals and 10 patients with adhesive capsulitis) underwent measurements of seven shoulder joint movements. Each movement was assessed three times by two raters using both iBalance and a goniometer, with measurements repeated after 1 week. The iBalance demonstrated excellent inter- and intra-rater reliability for flexion (ICC = 0.93 [0.91-0.95], 0.91 [0.88-0.94]), abduction (ICC = 0.97 [0.95-0.98], 0.93 [0.91-0.95]), and passive abduction (ICC = 0.97 [0.96-0.98], 0.98 [0.97-0.98]). The system also showed strong validity compared to the goniometer for flexion (ICC = 0.85 [0.68-0.92]), abduction (ICC = 0.95 [0.94-0.96]), and passive abduction (ICC = 0.97 [0.96-0.98]). Bland-Altman plots showed high consistency between the two devices for flexion, abduction, and passive abduction, with most data points falling within the limits of agreement. Patients with adhesive capsulitis exhibited greater variability than healthy individuals. No adverse events were reported, supporting the safety of the system. This study highlights the potential of a single-camera markerless motion capture system for diagnosing and treating shoulder joint disorders. The iBalance showed clinical applicability for measuring flexion, abduction, and passive abduction. Future enhancements to the algorithm and the incorporation of advanced metrics could improve its performance, facilitating broader clinical applications for diagnosing complex shoulder conditions.
Source: PubMed, quoted verbatim.
Cite
Lee S, Lee U, Kim Y, Noh S, Lee H, Lee S. Reliability and Validity of the Single-Camera Markerless Motion Capture System for Measuring Shoulder Range of Motion in Healthy Individuals and Patients with Adhesive Capsulitis: A Single-Center Study. Sensors (Basel). 2025. doi:10.3390/s25071960